Product Detail
Touch Screen Rubber And Plastic Weathering Aging UV Testing Machine
Product Applications:
UV accelerated weathering test chamber simulates damaging effects of long term outdoor exposure of sunlight, heat, moisture on materials and coatings. It exposes specimen to varying conditions of weathering- ultraviolet radiation, moisture and heat to test the performance and improve the quality.
Corresponding standard:
ASTM D4329, D499, D4587, D5208, G154, G53; ISO 4892-3, ISO 11507; EN534; prEN 1062-4, BS 2782; JIS D0205; SAE J2020
Technical Parameter:
- - - -
Model | KJ-9029 |
Test chamber size(LWH)MM | 1170450500 |
Outside dimension(LWH)MM | 13505401480 |
Chamber materials | 304# stainless steel |
Temperature range | RT+20~70 |
Temperature uniformity: | 1 |
Temperature fluctuation: | 0.5 |
Temperature control | PID SSR control |
Humidity range | 90%RH |
Controller | TEMI 880 programmable controller, touch screen, LCD display |
Control mode | Balance temperature humidity control (BTHC) |
Communication function | Be able to control the machine via computer using TEMI control software through RS-232 port on the machine |
Test cycle setting | Illumination, condensation and water spray test cycle is programmable |
Distance from specimen to lamp | 503mm (adjustable) |
Center distance between the lamps | 70mm | |
Lamp power & length | 40W/Piece, 1200mm/Piece | |
Lamps amount | 8 pieces of UVA-340 or UVB-313 imported lamps | |
Lifetime of lamp | 1600Hours | |
Irradiance | 1.0W/m2, adjustable directly on TEMI 880 touch screen controller | |
Wavelength of ultraviolet light | UVA is 315-400nm and UVB 280-315nm | |
Effective irradiation area | 900210mm | |
Irradiation black panel temperature | 50~70 | |
Standard specimen size | 75290mm/24 pieces | |
Water depth for water channel | 25mm, automatically control | |
Water consumption | 8L/day, tap water or distilled water | |
Testing time | 0999H, adjustable | |
Power | 1 phase, 220V/50Hz /10% 5KW | |
Protection | Overload short circuit protection, over temperature protection, water lacking protection | |
- - - -
Standard accessories:
1. One copy of English user guide.
2. One CD of controller software and operation instruction video.
Kindly note:
1. Please advise whether 8 UVA or 8 UVB lamps are needed when order.
2. UVA-340 lamps can well simulate the natural outdoor exposure. UVB-313 lamps are able to do accelerated test more quickly via stronger irradiation.
Safety protection:
1. Ground fault protection
2. Power overload, short circuit breaker
3. Control circuit overload, short circuit fuse
4. Water lacking protection
5. Over-temperature protection
Usage requirements:
1. Environmental temperature: 5~+32
2. Environmental humidity: 85% 3. Power requirement: AC220(10%)V/50HZ
4. Capacity: 5KW
5. Fine grounding
Pictures: